In Vivo Comparison of Air Charged Catheters with Water Filled Catheters for Intravesical and Intrarectal Pressures Recording during Urodynamic Testing. - Comparison AC-WFpressures
In Vivo Comparison of Air Charged Catheters With Water Filled Catheters for Intravesical and Intrarectal Pressures Recording During Urodynamic Study
Various systems to measure intravesical and intrarectal pressure during urodynamic testing; especially cystometry, exist. Water filled tube -systems are the most commonly used and should be regarded as the contemporary standard. A water filled system is however sensitive to tube and or patient movement artefacts and prone to erroneous calibration. Air charged catheters are less sensitive to patient and especially tubing- movements, and calibrate easier. However, in vitro tests have demonstrated that air charged catheters respond somewhat slower and relatively damped, especially to rapid pressure changes as in (simulated) coughing, in comparison with water filled systems. The clinical relevance of these observations is unknown.
This is a study to compare the technical reliability and clinical applicability of the two types of catheter systems for cystometry in a synchronous double catheter testing procedure in a prospective group or patients scheduled for urodynamic investigation.
100 项与 T-DOC Co. LLC 相关的临床结果
0 项与 T-DOC Co. LLC 相关的专利(医药)
100 项与 T-DOC Co. LLC 相关的药物交易
100 项与 T-DOC Co. LLC 相关的转化医学